Industrial ETO sterilizer

High-Volume ETO Sterilizer

Industrial ETO Sterilizer for High-Volume Medical Device Production

An industrial-scale EO gas sterilizer for controlled low-temperature processing of heat- and moisture-sensitive medical devices and products.

Chamber size customized to your needs
Double-door pass-through
Vacuum-based cycle operation
Designed according to ISO 11135

Interactive 3D

Explore the Industrial ETO Sterilizer

Drag to rotate, scroll or pinch to zoom, and tap the numbered points to see how each subsystem works.

Double hinged doors, radial locking

SS304 doors at both ends for pass-through loading. Each locks manually with a central handwheel driving radial locking bars. Interlocks allow only one door open at a time, and none until aeration is complete.

Loading 3D viewer…

Sterilizer Overview

EO gas sterilization for validated low-temperature loads

A double-door ethylene oxide sterilizer for vacuum (negative-pressure) cycles, with the chamber size customized to each customer's load requirements: a mirror-finish SS316 chamber with a hot-water jacket, manual radial-locking doors, water-ring vacuum pump, steam generator, EO vaporiser and weighing, a digital controller and SCADA batch reporting.

Who is it for?

Medical device manufacturers, contract sterilizers and pharmaceutical companies sterilizing large production loads.

What problem does it solve?

High-volume EO sterilization of packaged, heat- and moisture-sensitive products that cannot take steam, with pass-through loading that keeps the non-sterile and sterile sides apart.

Key features

Custom-sized industrial chamber

The chamber is sized to each customer's production volume and load requirements, built in 10 mm SS316 with a mirror finish.

Low-temperature EO gas sterilization

Designed for heat- and moisture-sensitive medical devices and products that are unsuitable for steam.

Touch-screen operation

Digital controller with a 10-inch colour touchscreen for cycle selection, status, alarms and guided operation.

Interlocked double doors

Door release is controlled by the system: only one door opens at a time, and only after aeration is complete.

Cycle records

SCADA software with batch-wise reports and three-level password access. Wi-Fi remote operation is included.

Integrated humidity and heating

An 18 kW steam generator supplies humidity, and a hot-water jacket heats the chamber evenly.

Cycle process overview

  1. 01

    Load and seal - Arrange packed products for gas penetration, load through the non-sterile-side door and close the chamber.

  2. 02

    Conditioning - Evacuate the chamber (down to −0.85 bar), then bring humidity and temperature to the programmed setpoint.

  3. 03

    EO exposure - Weighed EO is vaporised and injected, then held for the configured dwell time while 10 Pt-100 sensors log temperature.

  4. 04

    Evacuation - Remove EO gas from the chamber through controlled evacuation and flushing with 0.2 µm-filtered air.

  5. 05

    Aeration - Purge residual gas before the load is released. The doors stay locked until aeration is complete.

Safety and operation

  • Vacuum-based cycle operation (0 to −0.85 bar) designed for controlled EO exposure
  • Three-level pressure protection: safety valve, pressure switch with pneumatic release valve, and a digital pressure switch
  • Door interlocks: one door at a time, none until aeration is complete
  • EO gas monitor, status indicators, alarms and batch records for operator awareness

Technical snapshot

Method
Ethylene oxide gas sterilization
Chamber size
Customized to customer requirements
Cycle type
Vacuum / negative-pressure cycle only
Operating pressure
0 to −0.85 bar
Sterilant
100% EO or 90% EO + 10% CO₂
Doors
Double door, hinged, manual radial locking
Control
Digital controller, 10-inch colour touchscreen, SCADA

Applications

Where Industrial ETO fits

Industrial ETO is intended for facilities evaluating ethylene oxide sterilization where compatible loads require a controlled low-temperature EO gas process.

01

Medical device manufacturers

For high-volume EO sterilization of packaged disposables and devices on the production line.

02

Contract sterilization

For service providers sterilizing large, mixed loads for multiple customers.

03

Pharmaceutical companies

For packaging components and heat-sensitive materials.

04

Surgical dressings & textiles

For gauze, dressings, gowns and drapes.

Frequently asked questions

The chamber size is customized to each customer's production volume and load requirements. Every unit has doors at both ends so the load can pass from the non-sterile side to the sterile side. Contact SSQ to discuss the size you need.

Vacuum (negative-pressure) cycles only, with an operating range of 0 to −0.85 bar. Positive-pressure sterilization cycles are not supported.

The specification lists 100% EO or 90% EO + 10% CO₂. The final gas selection and system compatibility are confirmed with your order.

It depends on your products, load and validated process. Cycle time is established during cycle development and validation.

Related products

Need help evaluating Industrial ETO?

Share your facility type, load profile, chamber requirement, cycle needs, and documentation expectations. SSQ can help confirm whether this model or another platform is the right fit.

Request a Quote
Enquire